Product Details

SAB-500 Ophthalmic A/B Scanner

SAB-500

SAB-500 ophthalmic A/B scanner: 15-inch colour LCD on Windows. B, B+B, B+A and A modes; 10 MHz A-scan probe, 40 mm depth; IOL power calculation.

  • Brand: SonoStar
  • Price: On request
  • Probe frequency: 10 MHz (20 MHz probe optional); sector scan; multi-level continuous and real-time zoom; gain 30-105 dB
  • Connection: Self-contained 15-inch colour LCD, works standalone; A/B probes connect to the main unit by cable
  • Charging: Built-in battery, about 4 hours of operation (see specifications)
  • Category: Ophthalmic Ultrasound
Request a quote

Product overview

The SAB-500 is a standalone ophthalmic A/B scanner. It has three B-scan modes - normal, vitreous enhancement and retina observation - and is used to examine intraocular disease, showing where a lesion sits, how far it extends and how it relates to the surrounding tissue. Vitreous opacity, retinal detachment and intraorbital tumour are the usual findings. The A-scan measures anterior chamber depth, lens thickness and axial length, and the results are used to calculate intraocular lens power.

The whole system is built around a 15-inch colour LCD and a Windows operating system, so extra functions can be added later. A built-in battery gives about four hours of operation, which is enough to move between rooms without looking for a socket, and the unit measures 362 mm x 340 mm x 60 mm and weighs 2.5 kg.

The A-scan probe is a 10 MHz imported miniature probe that can be used handheld, in an immersion bath or mounted on a slit lamp, with contact and immersion scanning modes. Measurement modes cover phakic eyes, aphakic eyes and eyes with various intraocular lenses. Reports can be built from a custom template, and cases can be found again with multi-keyword search or with the preset expert dictionary.

Technical specifications

Specifications

Main unit

ItemData
Imaging modesB, B+B, B+A, A
Screen15-inch colour LCD
Findings promptPreset expert dictionary, or manual entry
SearchMulti-keyword search
ReportCustom report templates
Operating systemWindows, easy to extend
BatteryBuilt-in, about 4 hours of operation
Dimensions362 mm x 340 mm x 60 mm
Weight2.5 kg

A-scan

ItemData
Probe10 MHz imported miniature probe; handheld, immersion or slit-lamp mounting
Scanning modesContact, immersion
Detection depth40 mm
Eye modesPhakic, aphakic, eyes with various intraocular lenses
Measured valuesAnterior chamber depth, lens thickness, vitreous length - mean and standard deviation
Clinical accuracy+/-0.1 mm
Electronic accuracy0.0375 mm
IOL increment0.5 D
IOL formulasSRK-II, SRK-T, Binkhorst-II, Holladay, Hoffer-Q, Haigis (Standard)
Storage10 A-scan results

B-scan

ItemData
Probe frequency10 MHz (20 MHz probe optional)
ScanningSector scan
ZoomMulti-level continuous zoom, real-time zoom
Gain30-105 dB, adjustable
Detection depth60 mm
Scan angle53 degrees
Grey levels256
Pseudo-colourOCT pseudo-colour and several other maps
ResolutionAxial under 0.3 mm, lateral under 0.2 mm
MeasurementsMultiple distance, perimeter and area sets
EnhancementVitreous and retina enhancement
Image handlingImage saving and playback

Specifications updated 2026-11-01

Typical applications

The application scenarios on this page describe the device based on its technical data and configuration. The actual scope of application is the one stated in the product registration certificate. Follow your institution's procedures and the instructions for use.

Setting / departmentTypical useConfiguration note
OphthalmologyUltrasound examination of the vitreous, retina and other intraocular structuresDedicated ophthalmic probe
Ophthalmic biometryAnterior chamber depth, lens thickness and axial length measurementA-scan measurement
Ward and bedsideExamination away from a fixed workstationBuilt-in battery, about 4 hours

Usage and selection notes

The SAB-500 belongs to the ophthalmic series.

The actual scope of application is the one stated in the product registration certificate. Please confirm the model and configuration with our product department.

For contraindications and precautions, refer to the instructions for use. The device must be operated by a licensed physician with the relevant qualifications.

The technical data on this page is taken from the product technical documentation; where it has been updated, the latest version applies. The listed departments and application scenarios describe the device based on its technical data, and the actual scope of application is the one stated in the product registration certificate. This page makes no claim about clinical outcome and does not constitute clinical guidance.

Related

Related products

SAB-600 Ophthalmic A/B Scanner

SAB-600 ophthalmic A/B scanner...

View details →

SAB-200 Ophthalmic A/B Scanner

PC-based ophthalmic A/B scanne...

View details →

Need a quote or a sample for evaluation?

Tell us how you plan to use the system and our sales team will reply within one business day.

Get a quote
Home Call
Call +86 13434822215 (Mr. Zhong)